Phoenix x/ray nanotom series. Phoenix x|ray nanotom series cabinet x-ray systems are used in non-destructive testing, failure analysis, process control, and metrology.
FDA Recall Z-1828-2011 · Posted 2011-04-06
Advertisement
Recalling FirmGE Inspection Technologies, LP
Recall StatusTerminated
Product CodeRCE
Date Initiated2009-03-16
Date Terminated2017-04-17
Product Quantity118 units total
Reason for Recall
Some of the systems do not have a safety interlock on the service access panel. Although the systems have two independent means for indicating when x-rays are being generated, some systems may have one indicator that is not discernible. Also, some systems lack proper placement of a caution label and others do not have this label affixed.
Firm Action
The firm has completed all of the following necessary actions and no further actions have been indicated:
A Field Service Memo was issued to customers. It described the affected product and described the issue. GE Sensing and Inspection provided and installed retrofit kits, and the systems in question were tested by qualified Field Service Engineers at no charge to customers. GE asked customers to post the notification letter on the affected equipment until retrofit was completed. Questions were to be directed to the GE Service Center at 800-766-6062.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide Distribution