Terminated

Custom Cardiovascular Procedure Kits; Over-Pressure Safety Valves. For use only in the extracorporeal circuit for the cardiopulmonary bypass procedures for which the user designed it.

FDA Recall Z-1828-2015 · Posted 2015-06-23

Recall StatusTerminated
Product CodeDTL
Date Initiated2015-05-06
Date Terminated2015-07-27
Product Quantity163 devices

Reason for Recall

The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal blood loss, fluid weeping, or dripping on the user.

Firm Action

Consignees were notified of the recall by phone beginning on 5/7/2015 and a Urgent Medical Device Recall letter, dated 5/7/2015. The letter identified the affected device, reason for recall, potential hazard, and correction. Customers were asked to review the notice and ensure that all users are aware of the issue. In addition, they are to confirm the quantity of procedure kits for return; and, complete and return the Customer Response Form. Customers can call Terumo CVS Customer Service at 1-800-521-2818 with questions.

Root Cause

Component design/selection

Distribution Pattern

Distributed to the states of MO and FL.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑