EGIA 60 ARTICULATING EXTRA THICK SULU, Item code EGIA60AXT
FDA Recall Z-1828-2019 · Posted None
Advertisement
Recalling FirmCovidien LLC
Recall StatusOpen, Classified
Product CodeGDW
Date Initiated2019-05-17
Product Quantity3,113,280
Reason for Recall
The device may be missing one of two pin components that maintain alignment of the device jaws, resulting in incomplete staple formation which may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax which can result in the potential for infection and/or sepsis.
Firm Action
All consignees were notified by letter delivered via Federal Express and/or certified mail beginning May 17, 2019. The letter informs customers of the recall of specific item codes and production lots of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology due to the potential for a device to be missing one of two pin components that maintain alignment of the device jaws. Through this notification customers are instructed to take the following actions:
" Immediately quarantine and discontinue use of the affected item codes and lots.
" Return affected product.
" Forward the information to anyone who they have distributed affected Covidien Endo GIA" Articulating Reloads with Tri-Staple Technology.
" Complete the Recalled Product Return Form.
If you have any questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Root Cause
Process control
Distribution Pattern
Nationwide domestic distribution, worldwide foreign distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.