Open, Classified

CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.

FDA Recall Z-1828-2026 · Posted 2026-04-17

Recall StatusOpen, Classified
Product CodeLFL
Date Initiated2026-03-06
Product Quantity78 units

Reason for Recall

Software issue that renders the touch screen unresponsive.

Firm Action

An URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice dated 3/6/26 was sent to consignees. The notice instructs consignees to restart systems when freeze to restore functionality of the touchscreen capabilities for 8 minutes; this action may be used multiple times. Integra's Service and Repair can provide remote support to disable and re-enable the write filter with Integra employees and/or Authorized service personnel. This will reset the console to provide approximately 28 days with no touchscreen issue. Integra Lifesciences Service Team will contact consignees within 5 to 10 business days to schedule correction of affected devices at no cost. Consignees with any questions can direct them to Customer Service by emailing [email protected] or by phone at 1-800-654-2873.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide - US Nationwide distribution in the states of MI & NY. The countries of Belgium, China, Czechia, Dubai, Finland, France, Germany, Italy, Japan, Luxembourg, Netherlands. Poland, Spain, Switzerland, Tunisia & United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.