VIA Microcatheter VIA 21 REF VIA-21-154-01 - Product Usage: is a single lumen catheter designed to be introduced over a steerable guidewire into the vasculature.
FDA Recall Z-1829-2021 · Posted None
Advertisement
Recalling FirmMICROVENTION INC.
Recall StatusTerminated
Product CodeDQY
Date Initiated2021-04-21
Date Terminated2024-02-16
Reason for Recall
Due to inadvertently shipping to U.S. customers with international labeling which has some differences with US-cleared labeling.
Firm Action
On 04/30/ 2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Notification Letter via mail/hand delivery/phone to customers informing them that a limited-quantity of microcatheters with international labeling, which has some differences with US-cleared labeling, was inadvertently shipped to U.S. customers
Root Cause
Labeling Change Control
Distribution Pattern
U.S. Nationwide distribution in the states of FL, IA, MI, MN, TN and TX.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.