Coherence Dosimetrist, version 2.2. Intended use: Medical charged-particle radiation therapy system
FDA Recall Z-1834-2011 · Posted 2011-03-28
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeMUJ
Date Initiated2009-11-06
Date Terminated2011-04-06
Product Quantity60 units
Reason for Recall
A problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose, for the OAR (Organs at Risk) the user only defines the max. dose. The firms engineering team determined that the software defect was repeatable.
Firm Action
Affected consignees will be notified by the Customer Service Engineer to schedule a software update beginning November 2009.
Root Cause
Software design
Distribution Pattern
Worldwide distribution: USA including: CA, KY, MI, NJ and TN; and countries of: Canada and Mexico.
Related FDA 510(k) Records