Terminated

CIVCO Solstice(TM) SRS Immobilization System, Part Numbers: a) CHS03, b) CHS04 - Product Usage: The Tilting Head Fixation (Solstice) device is comprised of a baseplate and bowl with a locking lever. The bowl is permanently attached to the baseplate and the locking level allows the bowl to be tilted up to approximately 10 in relation to the baseplate at infinite intervals. The baseplate is attached to the treatment or simulation couch top or overlay by pins. When used with customizable cushion and thermoplastic mask, the system can immobilize the patient to under 1 millimeter. The provider can correct the patient position for simulation and treatments by tilting the device.

FDA Recall Z-1837-2019 · Posted None

Recalling FirmMed Tec Inc
Recall StatusTerminated
Product CodeIYE
Date Initiated2019-04-15
Date Terminated2020-10-05
Product Quantity28 units

Reason for Recall

There is a potential for movement of the Solstice Tilting Head Fixation System during patient setup and/or treatment.

Firm Action

The firm disseminated the notices by email on 04/15/2019 to their consignees. The notices stated the problem and provided mitigation strategies to employ to reduce the risk to the patient. The consignee was also provided the option to return the unit.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution - US Nationwide distributed in the states of CA, HI, PA, KY, FL, WI, AK and countries of Australia, Canada, Italy, Netherlands, Switzerland, Peru.