Ultraflex Tracheobronchial Stent System. Covered - 12/40/25 Catalog Number: M00569450. Single Use, Rx only. Boston Scientific, Natick, MA 01760. Indicated for the use in the treatment of tracheobronchial structures produced by malignant neoplasms.

FDA Recall Z-1842-2010 · Posted 2010-06-25

Recall StatusTerminated
Product CodeJCT
Date Initiated2010-05-17
Date Terminated2011-10-20

Reason for Recall

Suture related device failures during deployment and repositioning of the stent

Firm Action

Boston Scientific issued an "Urgent Medical Device Recall - Immediate Action Required" notification dated May 17, 2010 to U.S. Distributors stating that use and distribution of products should cease immediately and return all affected product to the firm for replacement. Distributors were asked to complete and return reply a verification tracking form. For further information, contact Boston Scientific at 1-866-868-4004.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - Only US distributed product is affected by recall