VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581

FDA Recall Z-1842-2014 · Posted 2014-06-20

Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeLPH
Date Initiated2014-05-19
Date Terminated2015-03-24
Product Quantity147 units

Reason for Recall

During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively.

Firm Action

On 6/2/2014 the firm sent "Urgent Medical Device Recall" letters to their customers. On 8/13/14, Zimmer contacted a distributor via phone pertaining to one additional lot (00-7843-011-06; lot 62547211) which had been included in the recall. Formal recall notices were issued to the distributor via email on 8/15/14.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, MA, MI, MN, NC, ND, NJ, NY, OH, OK, PA, TN, TX, VA & WI, and the countries of Canada, Colombia, Dominican Republic, Japan, Korea (S. Korea), Peru, Taiwan & Thailand.