Corin TriIFit TS SIZE 2, STANDARD TAPERED STEM, CEMENTLESS, REF 694.0002. Hip prosthesis component.
FDA Recall Z-1850-2021 · Posted None
Recalling FirmCorin Ltd
Recall StatusTerminated
Product CodeMEH
Date Initiated2021-05-06
Date Terminated2023-07-05
Product Quantity4 units
Reason for Recall
Units from one batch of TriFit TS size 2 stem were found to be labeled as a TriFit CF size 7 stem. And vice versa.
Firm Action
Corin initiated the recall by email on 05/06/2021 and followed with a letter on 05/07/2021 disseminated by email. The firm is seeking the return of the product.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the state of TX.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.