Pelton & Crane Spirit 1800 dental chair (Model #SP18)
FDA Recall Z-1851-2011 · Posted 2011-03-30
Advertisement
Recalling FirmPelton & Crane Company
Recall StatusTerminated
Product CodeNRU
Date Initiated2011-03-04
Date Terminated2011-08-16
Product Quantity6 units
Reason for Recall
The Link Arm Assembly that connects some rear mounted options to the Spirit 3000 or Spirit 1800 dental chairs may potentially break and fall off the dental chair.
Firm Action
Pelton & Crane sent an "Urgent Medical Device Recall" letter dated 3/04/2011. They were instructed to: Locate the affected devices this recall is in regards to. Schedule service with the end user for Pelton & Crane representative to replace the affected components this recall is in regards to. International accounts were instructed to order link arm assemblies from Pelton & Crane and replace the affected components. All consignees were to complete the Recall Acknowledgement/Return Form and return the form to Pelton & Crane.
Root Cause
Process control
Distribution Pattern
Distribution to USA, Canada, and the Netherlands.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.