Freedom EVO 200 - Product Usage: It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.
FDA Recall Z-1851-2019 · Posted None
Advertisement
Recalling FirmTecan US, Inc.
Recall StatusTerminated
Product CodeJQW
Date Initiated2019-04-12
Date Terminated2022-01-18
Product Quantity40
Reason for Recall
Gas springs not always replaced within the interval specified.
Firm Action
In April 2019, Tecan US sent Field Corrective Action Freedom EVO 200 letters to their US and Canadian consignees advising them of the issue; requesting that they complete and return an acknowledgement receipt; to pass the letter to any organization where the devices may be transferred and to arrange for a Tucan US representative to replace the gas springs
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide distribution. US Nationwide distribution in the states CA, CO, CT, KS, NC, NJ, OR, RI, TN, TX, and VA. Country of Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.