Product/Part: YelloPort Plus 5x70 Cannula + Luer/YC0507001. For use in laparoscopic procedures.

FDA Recall Z-1853-2022 · Posted 2022-09-29

Recall StatusOpen, Classified
Product CodeGCJ
Date Initiated2022-01-20
Product Quantity1

Reason for Recall

Laparoscopic devices have a sterile symbol included on labels but devices are non-sterile.

Firm Action

On 01/20/2022, Urgent Field Safety Notices were emailed to customers who were asked to do the following: 1) Quarantine affected devices and return to the recalling firm. 2) If the devices are no longer in stock, confirm whether devices were used clinically, and if so, whether the devices were sterilized in accordance with the Instructions for Use prior to use. 3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices were transferred. 4) Complete response form. The recalling firm provided the following email: [email protected]

Root Cause

Process control

Distribution Pattern

US: MA