Open, Classified

MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T

FDA Recall Z-1857-2022 · Posted 2022-09-29

Recall StatusOpen, Classified
Product CodeBSY
Date Initiated2022-08-03
Product Quantity136,080 units

Reason for Recall

Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.

Firm Action

On about 08/03/2022, Medline notified consignees via first class mail letter of the recall. The letter instructed consignees to immediately review inventory and destroy all affected product and complete the response form indicated at the link provided in the letter. Distributors were instructed to notify customers if further distributed, instruct those customers destroy any affected product, and include customer quantities on response form.

Root Cause

Under Investigation by firm

Distribution Pattern

Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Panama, Qatar, United Arab Emirates

Official Source

View FDA Recall →