Chest Support for R82 Mustang, Model Numbers: 861810, 861820, 861830; and Rabbit Up, Model Numbers: 869810, 869820, 869830
FDA Recall Z-1858-2020 · Posted None
Advertisement
Recalling FirmR82 A/S
Recall StatusTerminated
Product CodeITJ
Date Initiated2020-03-06
Date Terminated2023-09-05
Product Quantity2230 units
Reason for Recall
The plastic plug on the chest support for Mustang and Rabbit/Rabbit Up can come loose without use of a tool.
Firm Action
R82, Inc. notified their sole consignee on 02/09/2020 by email. The notice identified the choking hazard and requested that the consignee notify their customers of the hazard. The consignee will be sending letters beginning the week of 03/16/2020 to their customers.
Root Cause
Device Design
Distribution Pattern
NC