Open, Classified

Ultrasonic Cleaning Systems Triton 36 Model Number 6003000804 Triton 72 Model Number 6003000805 Ultrasonic cleaning system

FDA Recall Z-1859-2025 · Posted 2025-05-28

Recall StatusOpen, Classified
Product CodeFLG
Date Initiated2025-03-31
Product Quantity96 systems

Reason for Recall

During routine preventative maintenance checks, it was revealed that carbon filter sensors were breaking down due to electrochemical degradation.

Firm Action

On 03/31/2025, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via FedEx to customers informing them that as a result of an investigation, it has been revealed that a combination of chemical and electrical interactions that result in electrochemical degradation, compromising the structural integrity of the sensor over time. This degradation could potentially lead to water leakage into the machine. On 04/25/2025, the firm sent via FedEx and updated letter. Customers are notified and/or instructed that/to: -A field technicians will be contacting you to arrange for replacement if your unit has been impacted by this issue. -Complete and sign the attached response form to acknowledge that they have received the notification. Return the completed form to Getinge by emailing a scanned copy to [email protected] Public contact for questions/assistance - [email protected]

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide - U.S. Nationwide distribution including in the states of AZ, CA, CO, FL, ID, IL, IN, KS, MA, MD, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SD, TX, VA, WA, WI, and WY. The country of Canada.

Official Source

View FDA Recall →