Gambro Cartridge Blood Set, 0141080XXXC, is intended for single use in a hemodialysis treatment using the Phoenix and Centry system 3 Dialysis Delivery Systems.

FDA Recall Z-1861-2008 · Posted 2008-09-26

Recall StatusTerminated
Product CodeFJK
Date Initiated2008-05-02
Date Terminated2012-05-22

Reason for Recall

Dialysis tubing sets may have occlusions restricting blood flow.

Firm Action

Customers were sent an "Urgent Product Safety Notice" letter on 05/02/2008. They were instructed to visually inspect all blood sets prior to connection to the dialyzer to ensure there were no occlusions or partial occlusions. On 05/27/2008, an "Urgent Product Recall" letter was sent to all customers requesting the return of all remaining affected blood sets.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution --- including states of CA,GA, IL, KY, MO, NJ, NM, NY, NC, ND, PA, TN, and VA.