Triathlon Femoral Sizer - The Femoral sizer was developed for use with the Triathlon Total Knee Instrumentation system. It interfaces with the Modular Handled and the Femoral Stylus to facilitate sizing for a femoral component, and preparing holes to accept the corresponding 4:1 Cutting Guide. The femoral sizer is designed to prepare the femoral cuts in 3 degrees of external rotation by referencing against the posterior condyles. The Femoral sizer is set by locking it into the left or right orientation based on the femur.

FDA Recall Z-1864-2008 · Posted 2008-09-16

Recall StatusTerminated
Product CodeLXH
Date Initiated2005-05-27
Date Terminated2008-09-24
Product Quantity560 units

Reason for Recall

Stryker Orthopaedics became aware that when attempting to position the Triathlon Femoral Sizer, the pre-setting may slip from l (left knee) to R (right knee) and vice verse intraoperatively.

Firm Action

Letters were sent to all affected Stryker Branches/Agencies on May 27, 2005. Letters requested that each branch examine the inventory and hospitals locations to identify the instruments. Contact James Young at 201-831-5110, Brian Coughlin at 210-831-5665 or Rita Intorrella at 201-831-5825 if you have questions.

Root Cause

Device Design

Distribution Pattern

To Stryker branches/agencies worldwide.