Terminated

Encore Modular Femoral Neck, 12 degree/35 mm; Ref #410-35-112; distributed by Encore Medical, Austin, TX 78758. Designed to ensure the proper anteversion and femoral head orientation in the acetabulum.

FDA Recall Z-1868-2008 · Posted 2008-08-26

Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeJDI
Date Initiated2008-05-08
Date Terminated2009-09-08
Product Quantity20 units

Reason for Recall

Product not manufactured to specifications; modular femoral necks found to contain 24 scallops, instead of the standard 12.

Firm Action

Firm notified sales agents of recall on 05/08/08 via an Urgent: Medical Device Recall letter. Sales agents asked to immediately examine their inventory and quarantine product subject to recall. Agents are asked to notify their customers if they had further distributed the product. All product must be returned to Encore along with the response form.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution --- USA including states of IN, KS, GA, WA, SD, and countries of Greece and Australia.