Terminated

Accell Evo3C Demineralized Bone Matrix, 5 cc, Model Number: 02-6000-050 , Lot Number: 132160. Product Usage: Accell Evo3c Demineralized Bone Matrix Putty is intended for filling voids and gaps in the skeletal system that are not intrinsic to the stability of the bony structure. Accell Evo3c is indicated for use as a bone graft extender in the spine, extremities and pelvis, or as a bone void filler in the extremities and pelvis. The voids or gaps may be surgically created defects or the result of traumatic injury to the bone.

FDA Recall Z-1868-2014 · Posted 2014-06-25

Recall StatusTerminated
Product CodeMBP
Date Initiated2014-05-19
Date Terminated2014-09-05
Product Quantity61

Reason for Recall

This lot of Accell Evo3C Demineralized Bone Matrix, 5cc was incorrectly packaged in boxes that describe the product as Accell Evo3 Demineralized Bone Matrix. Evo3C is contraindicated for treatment of vertebral compression and not indicated for use in the posterolateral spine.

Firm Action

Integra sent an Urgent Voluntary Medical Device Recall dated May 19, 2014 to customers via mail (traceable courier service). The recall notification letter identified affected product, reason for recall and actions to be taken Customers were instructed to complete the "RECALL ACKNOWLEDGEMENT AND RETURN FORM". and return to: [email protected] or FAX to 1-609-275-9445. For questions contact Customer Service at 1-800-550-7155.

Root Cause

Packaging

Distribution Pattern

US Nationwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →