Terminated

DriSate Product Code DR-145 The acid concentrate powders and liquids are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/monitor and bicarbonate concentrate.

FDA Recall Z-1869-2018 · Posted None

Recalling FirmRockwell Medical, Inc
Recall StatusTerminated
Product CodeKPO
Date Initiated2018-04-07
Date Terminated2018-10-01
Product Quantity139

Reason for Recall

A subassembly component (dextrose bag) within this case product contained small amounts of sodium bicarbonate from a prior batch.

Firm Action

Removal notification involved one customer and two facilities. The two facilities receiving the DR-145 kit lot were contacted directly via phone, email, and issuance of a RGA to return the product. Stock retrieval was completed by 4/7/17. See attached email and RGA. Communication was already completed and stock was retrieved. Discard and/or return via RGA provided The complaint documented a stock reconciliation in the complaint and all units were accounted for. The two customer facilities involved responded directly Affected kits were reworked to remove and replace the nonconforming dextrose bag. For further questions, please call (864) 849-0530.

Root Cause

Employee error

Distribution Pattern

US Distribution to the state of NY