Shaver Handpiece Sterilization Tray. Model number 272-700-000. For sterilization of Stryker Endoscopy arthroscopic shaver.
FDA Recall Z-1872-2014 · Posted 2014-06-25
Advertisement
Recalling FirmStryker Endoscopy
Recall StatusTerminated
Product CodeHRX
Date Initiated2014-05-19
Date Terminated2015-05-07
Product Quantity1992 trays
Reason for Recall
STERRAD 100S parameters provided in the Shaver Handpiece reprocessing guide (P10299 Revision A) did not fully consider all worst-case scenarios during validation testing.
Firm Action
Device Correction letters were sent on May 23, 2014 to all users by certified mail. Internation sites will receive notification and be responsible for contacting customers.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.