Flexicair MC3 (model M4000) Low Airloss Therapy Units, for the prevention of pressure sores.
FDA Recall Z-1875-2009 · Posted 2009-08-03
Recalling FirmHill-Rom Manufacturing, Inc.
Recall StatusTerminated
Product CodeIOQ
Date Initiated2009-05-12
Date Terminated2012-04-03
Product Quantity889 units
Reason for Recall
The devices may emit smoke from the blower box assembly.
Firm Action
Urgent Field Safety Notice letter dated May 11, 2009. Hill-Rom service centers were notified by letter on 05/13/2009. Action to be taken are as follows: Rental Customers; In the next 60 days, Hill Rom will be removing the Flexicair II and Flexicair MC3 low Airloss Therapy Units from service to correct the devices. During that time, Hill-Rom may provide an equivalent substitute therapy to accommodate the patients who continue to need a pressure reducing therapy device. If t his malfunction occurs before a substitution or a corrected device is provided, unplug the device, remove the patient from the surface and notify Hill-Rom immediately. Facility-Owned Customers: In the next 60 days, for all facility owned Flexicair II and Flexicair MC3 Low Airloss Therapy Units, a Hill-Rom service technician will either replace the affected components in the blower assembly or deliver an equivalent substitute unit from Hill-Rom's rental fleet. If this malfunction occurs before the device is corrected or a substitute provided, do as follows: unplug the device, remove the patient from the surface, and immediately notify Hill-Rom Technical Support using the attached listing their geographical region.
Root Cause
Equipment maintenance
Distribution Pattern
Nationwide and Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.