Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes JUDKINS PACK DYNJ51126 ANGIO PACK DYNJ32555C, DYNJ66631 ANGIO TRAY DYNJ46153B ANGIOGRAPHY PACK DYNJ26855F ANGIOGRAPHY PACK-M-LF DYNJ24530I ANGIOGRAPHY TRAY-LF DYNJ0220880V, DYNJ30077D CATH LAB ANGIO PACK-LF DYNJ0545064X CATH LAB PACK DYNJ39057C FEMORAL ANGIOGRAPY SET UP DYNJ61015B RAD-ANGIO PACK DYNJ47710C SURGERY ANGIO PACK-LF PHS677784013A
FDA Recall Z-1876-2026 · Posted 2026-04-17
Advertisement
Recalling FirmMedline Industries, LP
Recall StatusOpen, Classified
Product CodeOEQ
Date Initiated2026-02-27
Product Quantity270,311 total
Reason for Recall
Unapproved design changes to the products outside of the 510(k) clearance.
Firm Action
On February 27, 2026, the firm began notifying customers via email and first-class mail.
Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Root Cause
Component change control
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.