PANTA Nails, 10mm x 150 length, Newdeal, Lyon France; Integra, 311 Enterprise Drive, Plainsboro, NJ. The Newdeal Ankle Nail is intended for use in tibiotalocalcaneal arthrodesis and the treatment of trauma to the hindfoot and distal tibia. The nail affords rigid, load sharing fixation that incorporates a simple nail-mounted, in-line method of compression across the arthrodesis site.

FDA Recall Z-1877-2008 · Posted 2008-09-16

Recall StatusTerminated
Product CodeHSB
Date Initiated2008-03-24
Date Terminated2008-09-16
Product Quantity45 units received in US, 22 were distributed.

Reason for Recall

Certain PANTA Nails from one lot were incorrectly manufactured. The internal thread for certain PANTA Nails (Part Number 500050ND; Lot Number: E7XX) may present an insufficient depth and may not allow the engagement of the threaded part of the compression device. The risks presented include: an increase in surgery time, or the implantation of a nail less ideally suited for the patient's anatomy

Firm Action

On March 24, 2008, Recall Notification Letters and Recall Acknowledgement Forms were sent to all Integra Sales Specialists to confirm products currently in their inventory and to place inventory on hold with no further shipments to customers. If you have any questions contact Judy O'Grady at (609) 936-2311.

Root Cause

Process control

Distribution Pattern

Class II Recall - Nationwide Distribution.