Boston Scientific, LeVeen Needle Electrode, 25 cm, 5.0 cm, sterile; REF M001262170. For tissue coagulation.

FDA Recall Z-1877-2009 · Posted 2009-09-09

Recall StatusTerminated
Product CodeGEI
Date Initiated2009-07-06
Date Terminated2010-10-08
Product Quantity347

Reason for Recall

The tines may detach from the core wire during use, which may result in bleeding or soft tissue trauma requiring medical intervention.

Firm Action

Boston Scientific notified consignees by letter dated 7/6/09 and instructed to cease using and to return the products, and to complete and return a reply form.

Root Cause

Process control

Distribution Pattern

Nationwide.