Terminated

Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.

FDA Recall Z-1877-2016 · Posted None

Recall StatusTerminated
Product CodeNJT
Date Initiated2016-04-05
Date Terminated2017-11-30
Product Quantity90 Units

Reason for Recall

Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling.

Firm Action

An Urgent: Medical Device Recall Notification dated 4/5/2016 was sent to consignees with the affected lots of CDX2 and RCC marker. The customers were notified of the problems as well as the following actions to be taken by the consignees: Do not use or continue to use the reagent lots listed above, as they have the potential to result in inadequate tissue staining and lead to weak or false negative pathology results. Appropriately destroy any unused or partially used affected lots of the reagent Complete the attached Recall Notification Acknowledgement Form that this action has been undertaken. The firm also request that the notification is passed to all those within their organization who need to be aware of this issue. The firm requests consignees to return the completed acknowledgement form to [email protected] to receive credit for affected lots that have been destroyed by the consignees laboratory.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Distributed in the following states AR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN and TX, and in Canada.