Terminated

Trident Endoscopic Ultrasonic Aspiration Needle (FNA)

FDA Recall Z-1878-2018 · Posted None

Recall StatusTerminated
Product CodeODG
Date Initiated2018-03-12
Date Terminated2018-08-30
Product Quantity40 total

Reason for Recall

The combination of the firmness of the mass, the potential endoscope and elevator position combined with the force required to penetrate the mass may contribute to the breakage of the needle in the human body, which could cause injury.

Firm Action

The firm emailed the distributor on March 12, 2018. (initial consignee). The consignee has been requested to return all products to the Manufacturer (Micro-Tech Nanjing).

Root Cause

Device Design

Distribution Pattern

US distribution to one distributor in MI.