Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes MINOR ENT SINUS PACK-LF DYNJ0101278F DYNJ0101278G DYNJ0101278I SINUS PACK DYNJ56262
FDA Recall Z-1880-2026 · Posted 2026-04-17
Recalling FirmMedline Industries, LP
Recall StatusOpen, Classified
Product CodeOGR
Date Initiated2026-02-27
Product Quantity270,311 total
Reason for Recall
Unapproved design changes to the products outside of the 510(k) clearance.
Firm Action
On February 27, 2026, the firm began notifying customers via email and first-class mail.
Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.
Root Cause
Component change control
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.
Official Source