MR Coaxial 10g x 140mm (VC10140MR).
FDA Recall Z-1881-2009 · Posted 2009-08-26
Advertisement
Recalling FirmBard Peripheral Vas
Recall StatusTerminated
Product CodeDWF
Date Initiated2009-03-26
Date Terminated2012-04-02
Product Quantity695
Reason for Recall
Some Vacora 10G Plastic Coaxial Cannulas may leave a small metal filament in the breast or the sample obtained post procedure.
Firm Action
Bard Peripheral Vascular, Inc. issued an "Urgent Medical Device Recall Notification" dated March 27, 2009 informing Customers of the affected devices. The firm requested that all affected product be removed and returned along with returning a Recall and Effectiveness Check Form via fax.
For further information, contact Bard Peripheral Vascular, Inc. at 1-800-321-4254.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.