Navitrack System - OS Knee Universal, Orthopedic Stereotaxic Instrument CAS Software application intended to assist in the placement of total knee replacement components

FDA Recall Z-1881-2017 · Posted 2017-04-26

Recall StatusTerminated
Product CodeOLO
Date Initiated2011-10-06
Date Terminated2017-05-08
Product Quantity105

Reason for Recall

Zimmer CAS voluntarily conducted a retrospective recall of the Navitrack System - OS Knee Universal software ORTHOsoft-UniTkr-2.3.2.6, due to a calibration sequence crash.

Firm Action

Zimmer sent a Medical Device Field Notification letter dated December 6, 2011, to all affected customers. This is a retrospective report of a correction taken on 6 December 2011. The software issues were fixed in the new software version 2.3.3.0. Recall notices were distributed on December 6, 2011. Telephone: 18663367846 (US & Canada) or 5143958883 (International), Email: [email protected] Customers with questions should call 1-866-336-7846 (US & Canada) 514-395-8883 (International). For questions regarding this recall call 574-372-4487.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - US including AZ CA CO FL GA IL IN KS MD MI MN MS OH OR PA SC SD TX VA WA and WI Internationally to Canada, Australia Austria Colombia Denmark France Germany Korea New Zealand Russia South Africa and Thailand