The Savaria Stairfriend is a Residential, commercial, indoor, stair lift for curved stairs.
FDA Recall Z-1882-2017 · Posted None
Advertisement
Recalling FirmSavaria Concord Lifts, Inc.
Recall StatusTerminated
Product CodeILK
Date Initiated2017-04-12
Date Terminated2017-10-16
Product Quantity226
Reason for Recall
Savaria Concord Lifts has decided to initiate a recall of the Stairfriend, because of product defect. The product defect relates to poor quality of welding holding the seat post .
Firm Action
The firm plans to notify its customers thru the phone and also via the dealers it sells it product too.
Root Cause
Nonconforming Material/Component
Distribution Pattern
North America / South America, please see attached list.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.