Terminated

The Savaria Stairfriend is a Residential, commercial, indoor, stair lift for curved stairs.

FDA Recall Z-1882-2017 · Posted None

Recall StatusTerminated
Product CodeILK
Date Initiated2017-04-12
Date Terminated2017-10-16
Product Quantity226

Reason for Recall

Savaria Concord Lifts has decided to initiate a recall of the Stairfriend, because of product defect. The product defect relates to poor quality of welding holding the seat post .

Firm Action

The firm plans to notify its customers thru the phone and also via the dealers it sells it product too.

Root Cause

Nonconforming Material/Component

Distribution Pattern

North America / South America, please see attached list.

Official Source

View FDA Recall →