Open, Classified

Cascade IOMAX Cortical Module, REF: 190296-200

FDA Recall Z-1882-2024 · Posted 2024-05-22

Recall StatusOpen, Classified
Product CodeGWF
Date Initiated2024-04-23
Product Quantity54 units

Reason for Recall

There is the potential that an electroneurodiagnostic medical device insulator may be installed incorrectly which could result in arcing to patient connections resulting in potential unintended shock to the patient.

Firm Action

On 4/22/24, Cadwell Industries, Inc. issued a "Urgent: Medical Device Recall" notification to affected consignees via Email: 1. Cease use of the device immediately and return all identified devices to Cadwell Industries, Inc. 2. Please complete the attached response form and return via email to [email protected]. 3. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred the IOMAX Cortical Module to. 4. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

Root Cause

Process design

Distribution Pattern

Worldwide - US Nationwide distribution in the state of MD and the countries of Mexico, Canada, Malaysia, India, Taiwan, Saudi Arabia, European Union, Switzerland, South Africa, Iraq, United Kingdom.