SmartSite Add-On Bag Access Device, Model No. 10013365
FDA Recall Z-1883-2017 · Posted None
Advertisement
Recalling FirmCareFusion 303, Inc.
Recall StatusTerminated
Product CodeFPA
Date Initiated2016-11-17
Date Terminated2018-10-29
Product Quantity384,923 units
Reason for Recall
CareFusion is recalling the SmartSite Add-On Bag Device due to reports of separation and/or leakages between the spike port and the drip chamber spike. A separation
or leak can cause a delay of infusion, an interruption of infusion, exposure to
medication or hazardous infusates, or underinfusion.
Firm Action
CareFusion sent an Urgent Medical Device Recall Notification letter dated November 2016, to inform their customers that CareFusion is recalling the SmartSite Add-On Bag Access Device Model 10013365 because of reports of leakages between the spike port and the drip chamber spike. The letter informs the customers of the lot numbers, potential risk and immediate actions to be taken. Customers with questions are instructed to contact CareFusion Support Center at (888) 562-6018 for recall related questions. Customers with adverse event report questions are instructed to contact Customer Advocacy at (888) 812-3266 or by email at [email protected]. Customers are instructed to complete and return the enclosed mandatory customer response card.
Root Cause
Other
Distribution Pattern
US Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.