Terminated

GE OEC 9800 Fluoroscopic X-Ray System, GE Healthcare, Surgery, Salt Lake City, Utah. The intended use of this device is to perform mobile fluoroscopic examination of human anatomy.

FDA Recall Z-1884-2008 · Posted 2008-09-25

Recall StatusTerminated
Product CodeJAA
Date Initiated2008-08-12
Date Terminated2012-04-23
Product Quantity201 systems

Reason for Recall

Beam limitation may be non-compliant on some X-ray units.

Firm Action

Consignees were notified by letter on 08/12/2008. GE service representatives will check proper beam compliance on next scheduled maintenance visit or will schedule a visit to address the issue. Information available at 1-800-874-7378.

Root Cause

Radiation Control for Health and Safety Act

Distribution Pattern

Nationwide