GE OEC 9900 Elite Fluoroscopic X-Ray System, GE Healthcare, Surgery, Salt Lake City, Utah. The intended use of this device is to perform mobile fluoroscopic examination of human anatomy.
FDA Recall Z-1885-2008 · Posted 2008-09-25
Advertisement
Recalling FirmOEC Medical Systems, Inc
Recall StatusTerminated
Product CodeJAA
Date Initiated2008-08-12
Date Terminated2012-04-23
Product Quantity335 systems
Reason for Recall
Beam limitation may be non-compliant on some X-ray units.
Firm Action
Consignees were notified by letter on 08/12/2008. GE service representatives will check proper beam compliance on next scheduled maintenance visit or will schedule a visit to address the issue. Information available at 1-800-874-7378.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide