EPi-Sense Guided Coagulation System with Visitrax, MOdel No. CDK-1413 - Product Usage: The EPi-Sense Guided Coagulation System with VisiTrax, consists of a sterile, single-use, disposable coagulation electrode device (1cm, 2cm, and 3cm sizes provided) intended to be used to coagulate cardiac tissue. The flexible, cooled electrode device, with a suction stabilizer feature, transmits radiofrequency (RF) energy from an Electrosurgical Generator (non-sterile, re-useable) connected through an Instrument Cable (sterile). A temporary sensing electrode feature may be used with an off-the-shelf electrogram recording system with the use of an additional instrument cable (non-sterile). An accessory Cannula may be used to facilitate coagulation device access and visibility of the heart. The EPi-Sense Guided Coagulation System with VisiTrax technology is intended for the coagulation of cardiac tissue using Radiofrequency (RF) energy using thoracoscopic, endoscopic, and laparoscopic surgical techniques and may be used for temporary cardiac signal sensing and recording during surgery when connected to a temporary external recording device.

FDA Recall Z-1887-2019 · Posted None

Recalling FirmAtriCure, Inc.
Recall StatusTerminated
Product CodeOCL
Date Initiated2019-05-29
Date Terminated2020-05-04
Product Quantity497

Reason for Recall

The sterile package seal for the Epi-Sense Guided Coagulation System may be compromised, and has a worst-case reasonable harm of infection if used.

Firm Action

On May 29, 2019, the firm notified customers of the recall via Urgent Advisory Notice. Customers were asked to take the following actions: " Immediately quarantine any affected product as identified within this Advisory Notice. " Contact AtriCure product complaints by phone at 1-866-349-2342 (select option 6) or e-mail to [email protected] to request a Return Goods Authorization (RGA). " Return the attached Acknowledgement Form. See Attachment B. The acknowledgement form must be completed and returned even if product is not on hand by selecting the second option. " Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. If you have any questions, please contact Rob Cantu, Vice President of Quality at (1-513-644-4245) from 9-6pm ET on Mondays - Fridays. You may also contact customer service at (1-866-349-2342) any time of day, your message will be forwarded to Quality Assurance for review promptly. This advisory issue will also be posted on AtriCures website at www.atricure.com/products.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - US Nationwide distribution, and countries of Germany and Netherlands.