Inpeco FlexLab Centrifuge Module tube loading algorithm, Inpeco SA P/N: FLX-202, with Firmware Versions: CM_1-7-0.H86 or xCMd_2-5-0.elf and CM_1-7-1.H86 or xCMd_2-5-1-849fad9.elf

FDA Recall Z-1887-2020 · Posted None

Recalling FirmInpeco S.A.
Recall StatusOpen, Classified
Product CodeJJE
Date Initiated2020-04-07
Product Quantity65 units

Reason for Recall

The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.

Firm Action

The firm sent an Urgent Medical Device Correction letter to all affected customers on March 30, 2020. The letter instructed the customers to complete and return the Customer Letter Receipt Confirmation and Implementation Check form. The firm explained that their regular service provider will contact them to schedule the firmware version upgrade.

Root Cause

Software design (manufacturing process)

Distribution Pattern

Worldwide distribution including the following states: Ohio, New York, Virginia, Pennsylvania, Texas, Delaware, California, Florida, Kentucky, Tennessee, Maryland, Rhode Island, Alabama, Massachusetts, Connecticut, Minnesota, Alabama, Montana, Louisiana, New Jersey, Virginia, New Hampshire, Michigan, North Carolina, Washington, Iowa, New Mexico, Nebraska, and Missouri