Open, Classified

Medline Convenience Kits containing select SKUs of 10mL Polycarbonate Colored Syringes EXTREMITY PACK DYNJ34846C HAND PACK DYNJ83218A, DYNJ85396 LAMINECTOMY DYNJ905156R MAJOR EXTREMITY DYNJ905159L, DYNJ905159M MAJOR EXTREMITY PACK DYNJ69926A PODIATRY PACK DYNJ41250B

FDA Recall Z-1887-2026 · Posted 2026-04-17

Recall StatusOpen, Classified
Product CodeOJH
Date Initiated2026-02-27
Product Quantity270,311 total

Reason for Recall

Unapproved design changes to the products outside of the 510(k) clearance.

Firm Action

On February 27, 2026, the firm began notifying customers via email and first-class mail. Customers were instructed to request over-labels to place on affected kits containing recalled syringes instructing them to remove the recalled component and replace with product from their supply.

Root Cause

Component change control

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Panama, Barbados.