Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only. Used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature.

FDA Recall Z-1891-2018 · Posted 2018-05-17

Recalling FirmVolcano Corporation
Recall StatusTerminated
Product CodeIYO
Date Initiated2018-03-22
Date Terminated2021-04-28
Product Quantity8 devices

Reason for Recall

The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.

Firm Action

Customers were notified on 03/21/2018. U.S. customers were notified via hand-delivered Urgent Medical Device Recall letters; foreign customer letter was mailed. The letter identified the affected device and the reason for the recall. Instructions included to stop usage of the device and to call remote support at 1-800-228-4728, to schedule a visit to replace the affected device. Customers were instructed to complete and return the customer response form or to provide the form to Philips representatives. Customers were also instructed to provide copies of the recall notification to all personnel within their organization who handled this device.

Root Cause

Other

Distribution Pattern

Worldwide Distribution -- U.S., to the states of California, Connecticut, Tennessee, North Carolina, Maryland, and Oregon; and, the country of Australia.