Edwards Lifesciences LLC, Fem-Flex II Femoral Venous Cannula, 12 Fr. x 11.5 cm, REF FEMII012V, Cardiopulmonary bypass tubing.
FDA Recall Z-1893-2008 · Posted 2008-09-11
Advertisement
Recalling FirmEdwards Lifesciences Research Medical, Inc.
Recall StatusTerminated
Product CodeDWF
Date Initiated2008-02-25
Date Terminated2008-09-16
Product Quantity25 units
Reason for Recall
Femoral Venous Cannula was color coded as an arterial cannula. This problem could result in a minor delay in the procedure. In the worst case, this product could be connected incorrectly during bypass, which would result in a more significant delay and has the potential to result in a permanent injury.
Firm Action
Consignees were notified by Edwards Lifesciences sales representatives and a Product Recall letter dated 2/22/08. Users were instructed to check their inventory for any affected product and to return all unused product to the recalling firm. Users were also requested to return a recall response form indicating the quantity of affected product.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution.