Signature Laparoscopic Instruments, Sharp Tooth Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5863020.

FDA Recall Z-1896-2025 · Posted 2025-06-02

Recalling FirmSilex Medical, LLC
Recall StatusOpen, Classified
Product CodeGEI
Date Initiated2024-01-09
Product Quantity29 units

Reason for Recall

Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

Firm Action

Silex formally notified customers on 01/09/2024. Instructions to consignees included to inspect all instruments that they received from the affected lots for the defect. If any instruments had the issue, the customers were instructed to send it back to their facility under an RGA.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.