Terminated

Adult Divided Cannula end tidal C)2 sampling w/male luer-lock connector. 7 ft, oxygen tube, 7 ft. CO2 tube RX only.For Single Patient Use only. Distributed by Tri-Anium Assembled in Mexico. Usage: Delivery of Anesthesia.

FDA Recall Z-1898-2012 · Posted 2012-06-27

Recalling FirmSALTER LABS
Recall StatusTerminated
Product CodeCCK
Date Initiated2012-05-30
Date Terminated2012-07-27
Product Quantity242 cases

Reason for Recall

A complaint was received, and later verified, that product labeled as SO-1296 with male luer lock connector actually contained product with a female luer lock connector.

Firm Action

The firm, Salter Labs, called their customers to discuss the recall and followed up with a "PRODUCT RECALL" letter dated May 29, 2012. The letter was distributed to all customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine their inventory and return any of the lots in their possession to Salter Labs; contact Salter Labs Customer Service at 1-800-235-4203 to arrange for return of the product, and complete and return the attached form/record via fax to: 661-854-6816 or Toll Free 1-800-628-4690. Should you have any questions, please contact Salter Labs at 1-800-235-4203.

Root Cause

Packaging

Distribution Pattern

Nationwide distribution: USA including states of: FL, IN, KS, PA and TX.