Terminated

Vertier Surgical Table, an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system. Part Numbers: 0788100000 Standard table, 0788100001 Standard table with 5th wheel, 0788100000S SIDNE-enabled standard table, and 0788100001S SIDNE-enabled 5th wheel table

FDA Recall Z-1899-2009 · Posted 2009-08-26

Recall StatusTerminated
Product CodeFQO
Date Initiated2008-11-13
Date Terminated2009-08-28
Product Quantity31

Reason for Recall

Surgical tables were produced using an unapproved tilt frame bushing.

Firm Action

Stryker Communications issued an "Urgent: Medical Device Recall" notice dated November 13, 2008 informing Consignees of the affected product. The firm intends to contact each facility to coordinate service by a trained representative. For further information, contact Stryker at 1-972-410-7100.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution -- United States (OR, IL, WA, CA, AR, TX, UT, SC and GA) and Canada.