Terminated

Sorin Stockert Heater-Cooler System 3T, a) Item #: 16-02-81, 240 v/60 Hz; b) Item #: 16-02-82, 208 v/60 Hz; c) Item #: 16-02-85, 120 v/60 Hz

FDA Recall Z-1901-2018 · Posted 2018-05-18

Recalling FirmLivaNova USA
Recall StatusTerminated
Product CodeDWC
Date Initiated2018-04-20
Date Terminated2025-02-05
Product Quantity2408

Reason for Recall

Firm is notifying customers that the Deep-Cleaning Service is now available in the United States. Stockert Heater-Cooler System 3T devices are suspected of microbial contamination and may contain visible biofilm.

Firm Action

On 04/20/2018, a Medical Device Correction notice was sent to customers informing them to return devices less than ten years old. Returned devices will receive a deep cleaning service, which will reduce microbial contamination to HPC < 100 CFU/ml of water and NTM < 1 CFU/100ml.

Root Cause

Device Design

Distribution Pattern

MN, TX, WA, OR, IL, OH, SC, AK, PA, CA, WI, GA, NJ, FL, AZ, KY, MS, MO, LA, AR, MA, MI, MT, CO, TN, NE, NC, VA, NV, HI, PR, WY, AL, OK, IN, CT, ME, IA, SC, KS, NM, SD, DC, MD, ND,