VARIANT II TURBO Hemoglobin Testing System with CDM 3.6T Software Version 3.5; Model Number: 270-2600; Product is manufactured and distributed by Bio-Rad Laboratories, Diagnostics Group, Hercules, CA. Provides an integrated method for sample preparation, separation and the percent determination of percent hemoglobin A1c in EDTA human whole blood.

FDA Recall Z-1902-2010 · Posted 2010-06-25

Recall StatusTerminated
Product CodeLCP
Date Initiated2009-01-22
Date Terminated2011-03-22

Reason for Recall

Sample test results for the first run appears to be correct but all subsequent results during that run were assigned incorrect barcode and tube position.

Firm Action

Bio-Rad Laboratories, Inc. issued and "Urgent: Medical Device Correction" dated January 22, 2009 to all affected consignees by FED EX, followed up by telephone calls for those not responding. Consignees were instructed on proper steps to take in dealing with the affected product. Further communication will follow to provide a solution. For further information, contact Bio-Rad Laboratories at 1-510-724-7000.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- United States, Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Czech Republic, France, Germany, Hong Kong, Hungary, India, Italy, Mexico, New Zealand, Poland, Singapore, South Korea, Sweden, Taiwan, Thailand and the United Kingdom.