smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity levels are compatible with an adequate long-term result; failed osteotomies, unicompartmental replacement, or total knee replacement. The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e., medial collateral and/or lateral collateral ligament) are absent or incompetent.
FDA Recall Z-1902-2018 · Posted None
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusTerminated
Product CodeMBH
Date Initiated2018-04-09
Date Terminated2019-02-22
Product Quantity18 units
Reason for Recall
A single lot of GENESIS II FEMORAL FIVE DEGREE VALGUS BUSHING due to a laser mark labeling error. The left valgus bushing was incorrectly laser marked as right and vice versa.
Firm Action
The firm sent an Urgent Medical Device Recall Notice dated April 9, 2018.
Required Actions:
1. Please inspect your inventory and locate any unused devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately.
a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out.
2. If you have no product to return, please put an X in the appropriate location below.
3. If you have product to return, please list the batch numbers and quantities of each batch that you are returning in the appropriate boxes below.
4. Complete the remainders of the form sign and send to [email protected] or fax to 901-566- 7975.
Please Note even if you have no product to return, this form must be completed, signed and returned.
5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number.
If you have any questions or concerns regarding this recall please contact [email protected]. For further questions, please call (978) 749-1440.
Root Cause
Process control
Distribution Pattern
US Distribution to the states of : NC, IN, CO MI, PA, NY, MA, CA, GA
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.