Novation Press-Fit Tapered, Standard Offset, 12/14 Taper, Size 14, Femoral Stem, Catalog Number 160-00-14; For use in primary total hip arthroplasty.
FDA Recall Z-1903-2016 · Posted None
Advertisement
Recalling FirmExactech, Inc.
Recall StatusTerminated
Product CodeMRA
Date Initiated2016-04-12
Date Terminated2019-07-24
Product Quantity1 device.
Reason for Recall
Devices were inadvertently sent to packaging and finished goods prior to the PMA site-change supplement approval.
Firm Action
Customers were notified 4/12/2016 with a Product Recall Advisory Notice of affected products.In order to comply with applicable regulations and Exactech policies: " Immediately cease distribution or use of these products. " Extend this information to your accounts that may have this product in their possession. " Identify and quarantine any of the subject devices in your inventory " Complete and Return the attached Recall Inventory Response Form to Exactech, per the instructions provided on the notice itself. According to policy, please contact your Exactech inventory representative within 5 business days to confirm quantities at your location. If you have any inventory restocking questions related to this issue, please contact Kaya Davis at [email protected] or 1-800-392-2832. Please return the products to the following address:
Exactech Distribution Center 2411 NW 66th Court Gainesville, Florida 32653
Attention: Kaya Davis [email protected] 352-337-3915 (Fax)
Root Cause
Packaging process control
Distribution Pattern
Distributed in FL, NJ, NY, OH, OK, RI, TN, VA and HI.