Siemens Artis Q.zen ceiling-Interventional Fluoroscopic X-Ray System Model Number: 10848354

FDA Recall Z-1903-2019 · Posted None

Recall StatusTerminated
Product CodeOWB
Date Initiated2019-06-06
Date Terminated2021-02-09
Product Quantity15

Reason for Recall

Artis Zee, Artis Q, Artis Pheno or Artis Icono systems equipped with a DCS Extended or DCS Movable, the potential exists for individual screws of the screwed connection joining the pivot bearing to the ceiling-mounted support to work loose or snap off, can result in severe injuries to patients and operating personnel

Firm Action

Siemens issued letter dated 6/6/19 advising of the problem, health risk and action to take: .Siemens is implementing a corrective action to all potentially affected users via Update Instruction AX023/18/S. Siemens Service organization will replace the screwed connection of the ceiling support according to specifications / instructions. In the event the bolt connection cannot be replaced due to technical conditions, the complete bearing will be replaced.

Root Cause

Device Design

Distribution Pattern

Nationwide