MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

FDA Recall Z-1905-2025 · Posted 2025-06-04

Recall StatusCompleted
Product CodeSAB
Date Initiated2025-04-23
Product Quantity3 units

Reason for Recall

The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.

Firm Action

A "MEDICAL DEVICE CORRECTION NOTIFICATION" letter dated 4/22/25 was sent to customers. Actions to be taken by the Customer/User: Do not use the device and transition scheduled procedures to another minimally invasive surgical technique i.e. laparoscopic or other robotic assisted device. Remove the device from service using your internal institutional procedures (e.g. Lock-Out, Tag-Out) for device maintenance until Virtual Incision technicians service the devices on hand. There is no anticipated impact to the conduct of the clinical study under the IDE. Please complete the attached Acknowledgement and Receipt form and return to Virtual Incision to indicate you received this letter. Customer Service, Clinical Support Staff, [email protected],913-608-3214. Quality Assurance, Sr Director of Quality and Manufacturing Engineering, [email protected], 308-529-1987. Regulatory Affair, VP of Clinical and Regulatory Affairs, [email protected], 352-275-726. Type of Action by the Company: Virtual Incision will be working with your site representatives to schedule on-site servicing of the device with updated software for the Surgeon Control Console within 30 days from receipt of this letter. Immediate corrections have been implemented through the analysis of the system data. Results indicate that implemented software updates have resolved the issue that could contribute to this deficiency in the future. If you have any question, call 513-500-4459 or email: [email protected].

Root Cause

Software design

Distribution Pattern

US Distribution to States: CA, FL, NE

Official Source